Analyseurs

BM 200 2nd Generation

  • Throughput - 240 tests/hour (without ISE), 360 tests/hour
    (with ISE Na, K, Cl)
  • Optional build-in PC
  • Special anti-noise and high ergonomic design
  • Methodology - end-point, two-point and kinetic,
    colorimetric and turbidimetric measurement
  • QC - Westgard, Multi rules and L-J rules
  • Conection to LIS - HL7 Protocol
  • Barcode reader for sample and reagents
  • Water consumption - max 2,5 l/hour
  • Optional ISE module
  • Optional Touch screen
  • 40 sample positions - no special positions for calibrators,
    controls and stat samples-possibility of use as many controls
    and calibrators as needed
  • Up to 5 virtual sample trays
  • 40 reagent positions, possible input 2 lots of the same reagent
  • or share reagent between tests
  • Possible reagents recognition by barcode reader
  • Refrigerated sample/reagent compartment 4-10°C
  • Real time reagent volume monitoring
  • 81 reaction cuvettes with special ultraviolet transmitting material
  • Hot air thermostat
  • Full-sealed matrix spectrometric system
  • 8 wavelengths from 340-670 nm, 800 nm optional
  • Halogen lamp with 2000 working hours life
  • Sample and reagent probe with liquid level detection
    and 3-D collision protection.
  • Maintenance free dilutor
  • Minimum reaction volume - 150 μl
  • Contamination free sample probe and stirrer

BM 300

  • Throughput - 300 tests/hour (without ISE), 420 tests/hour
    (with ISE Na, K, Cl)
  • Optional: ISE module, Internal PC with touch screen, barcode reader
  • High ergonomic design (special place for reagents screw
    and reaction cover)
  • Cuvette washing system: 6-step wash station with washing solution
  • 81 high light transmittance reaction cuvettes, optional quartz
    glass cuvettes
  • Anti-contamination design (automatic probe and mixer washing,
    intelligent anti-contamination program, carry-over contamination
    rate ≤ 0.005%)
  • Anti-noise design (silent water cooling, operating noise < 65dB)
  • Sample and Reagent tray – 40 positions for reagent, 40 positions
    for sample with STAT priority
  • Refrigerated sample/reagent compartment 4-10°C
  • Hot air thermostat for reaction cuvette
  • Minimum reaction volume - 150 μl
  • Longest reaction time – 15 min (single reagent); 12 min (double reagents)
  • Test method -Endpoint, Two-point, Kinetic, Colorimetric,Turbidimetric
  • Cal method - Linear (single-point, two- and multi-point), Logit-4P,
    Logit-5P, spline, exponential, polynominal
  • Wavelengths (nm) - 340, 405, 450, 510, 546, 578, 630, 670, optional 800
  • Halogen-tungsten lamp with > 2000h life time
  • Overall water consumption : < 6,5 l/hour;
  • User friendly software (multi-language; easy operating with iconic
    interface; powerful statistic system; manual dilution; auto dilution
    and post dilution; easy rerun and restore test)
  • Calibration and QC (Multiple calibration module, Advanced algorithm to obtain
    best calibration curve; Multi QC rules: Westgard, Cumulative sum check, Twin plot)
  • Real Time Monitoring (reaction, calibration and QC test curve; sample, reagent
    and reaction cuvette status; lamp intensity, distilled water and waste tank;
    reaction and reagent disk temperature).

BM 500

  • Throughput – up to 400 tests/hour (560 with ISE)
  • ISE module for Na, K, Cl as option
  • Build-in barcode reader for samples and reagents
  • Reaction cuvettes – reusable 100 reaction cuvettes
  • 8 steps cuvettes washing system
  • Wavelengths – 12 wavelengths in the range 340 to 800 nm
  • Two arm for mixing of sample and reagents
  • Sample probe - with liquid level detection and collision protection
  • and clot detector
  • The volume of dispensed sample - 2-45 μl
  • Up to 4 reagents in one reaction
  • The volume of reagent dispensed - R1, R3 - 150-350 μl,
  • R2, R4 - 20-250 μl
  • Positions for reagents – 80 bottles 20, 30 ml and 70 ml,
  • cooled to 2-10°C
  • Positions for samples – 100 tubes with diameter 12 - 14 mm
  • height 35-100 mm, microcups
  • Connection to LIS – HL7 protocol

 

BM 100

  • Throughput – up to 150 tests/hour (560 with ISE)
  • Build-in barcode reader for samples and reagents
  • Reaction cuvettes – reusable 81 reaction cuvettes
  • Wavelengths – 8 wavelengths in the range 340 to 670 or 800 nm
  • Additional arm for mixing of sample and reagents
  • Sample probe - with liquid level detection and collision protection
  • and clot detector
  • The volume of dispensed sample - 2-30 μl
  • The volume of reagent dispensed - R1: 150–300 μl; - R2: 20–300 μl
  • Reaction volume - 150 - 500 μl
  • Positions for reagents – 40 bottles 20, 30 ml, cooled to 4-10°C
  • Positions for samples – 40 tubes with diameter 12 - 14 mm, height 35-100 mm, microcups
  • Water consumption 1,5 L / hour.